In the deuterium atom ( 2H ), with one unpaired electron, one unpaired proton, and one unpaired neutron, the total electronic spin = 1/2 and the total nuclear spin = 1.
Two or more particles with spins having opposite signs can pair up to eliminate the observable manifestations of spin. An example is helium.
In nuclear magnetic resonance, it is unpaired nuclear spins that are of importance.
For hydrogen, γ = 42.58 MHz / T.
Nuclei | Unpaired Protons | Unpaired Neutrons | Net Spin | γ (MHz/T) |
---|---|---|---|---|
1H | 1 | 0 | 1/2 | 42.58 |
2H | 1 | 1 | 1 | 6.54 |
31P | 1 | 0 | 1/2 | 17.25 |
23Na | 1 | 2 | 3/2 | 11.27 |
14N | 1 | 1 | 1 | 3.08 |
13C | 0 | 1 | 1/2 | 10.71 |
19F | 1 | 0 | 1/2 | 40.08 |
When the proton is placed in an external magnetic field, the spin vector of the particle aligns itself with the external field, just like a magnet would. There is a low energy configuration or state where the poles are aligned N-S-N-S
and a high energy state N-N-S-S.
In NMR and MRI, the quantity ν is called the resonance frequency and the Larmor frequency.
When the energy of the photon matches the energy difference between the two spin states an absorption of energy occurs.
In the NMR experiment, the frequency of the photon is in the radio frequency (RF) range. In NMR spectroscopy, ν is between 60 and 800 MHz for hydrogen nuclei. In clinical MRI, ν is typically between 15 and 80 MHz for hydrogen imaging.
The CW experiment can also be performed with a constant magnetic field and a frequency which is varied. The magnitude of the constant magnetic field is represented by the position of the vertical blue line in the energy level diagram.
At room temperature, the number of spins in the lower energy level, N+, slightly outnumbers the number in the upper level, N-. Boltzmann statistics tells us that
E is the energy difference between the spin states; k is Boltzmann's constant, 1.3805x10-23 J/Kelvin; and T is the temperature in Kelvin.
As the temperature decreases, so does the ratio N- /N+. As the temperature increases, the ratio approaches one.
The signal in NMR spectroscopy results from the difference between the energy absorbed by the spins which make a transition from the lower energy state to the higher energy state, and the energy emitted by the spins which simultaneously make a transition from the higher energy state to the lower energy state. The signal is thus proportional to the population difference between the states. NMR is a rather sensitive spectroscopy since it is capable of detecting these very small population differences. It is the resonance, or exchange of energy at a specific frequency between the spins and the spectrometer, which gives NMR its sensitivity.
At any instant in time, the magnetic field due to the spins in each spin packet can be represented by a magnetization vector.
The size of each vector is proportional to (N+ - N-).
The vector sum of the magnetization vectors from all of the spin packets is the net magnetization. In order to describe pulsed NMR is necessary from here on to talk in terms of the net magnetization.
Adapting the conventional NMR coordinate system, the external magnetic field and the net magnetization vector at equilibrium are both along the Z axis.
It is possible to change the net magnetization by exposing the nuclear spin system to energy of a frequency equal to the energy difference between the spin states. If enough energy is put into the system, it is possible to saturate the spin system and make MZ=0.
The time constant which describes how MZ returns to its equilibrium value is called the spin lattice relaxation time (T1). The equation governing this behavior as a function of the time t after its displacement is:
T1 is therefore defined as the time required to change the Z component of magnetization by a factor of e.
If the net magnetization is placed along the -Z axis, it will gradually return to its equilibrium position along the +Z axis at a rate governed by T1.
The equation governing this behavior as a function of the time t after its displacement is:
The spin-lattice relaxation time (T1) is the time to reduce the difference between the longitudinal magnetization (MZ) and its equilibrium value by a factor of e.
The time constant which describes the return to equilibrium of the transverse magnetization, MXY, is called the spin-spin relaxation time, T2.
T2 is always less than or equal to T1.
The net magnetization in the XY plane goes to zero and then the longitudinal magnetization grows in until we have
Mo along Z.
Any transverse magnetization behaves the same way.
The transverse component rotates about the direction of applied magnetization and dephases. T1 governs the rate of recovery of the longitudinal magnetization.
In summary, the spin-spin relaxation time, T2, is the time to reduce the transverse magnetization by a factor of e. In the previous sequence, T2 and T1 processes are shown separately for clarity. That is, the magnetization vectors are shown filling the XY plane completely before growing back up along the Z axis. Actually, both processes occur simultaneously with the only restriction being that T2 is less than or equal to T1.
Two factors contribute to the decay of transverse magnetization.
1) molecular interactions (said to lead to a pure pure T2 molecular effect)
2) variations in Bo (said to lead to an inhomogeneous
T2 effect
The combination of these two factors is what actually results in the decay of transverse magnetization. The combined time constant is called T2 star and is given the symbol T2*. The relationship between the T2 from molecular processes and that from inhomogeneities in the magnetic field is as follows.
A magnetization vector rotating at the Larmor frequency in the laboratory frame appears stationary in a frame of reference rotating about the Z axis. In the rotating frame, relaxation of MZ magnetization to its equilibrium value looks the same as it did in the laboratory frame.
A transverse magnetization vector rotating about the Z axis at the same velocity as the rotating frame will appear stationary in the rotating frame.
A magnetization vector traveling faster than the rotating frame rotates clockwise about the Z axis.
A magnetization vector traveling slower than the rotating frame rotates counter-clockwise about the Z axis .
In a sample there are spin packets traveling faster and slower than the rotating frame. As a consequence, when the mean frequency of the sample is equal to the rotating frame, the dephasing of MX'Y' looks like this.
In a frame of reference rotating about the Z axis at a frequency equal to that of the alternating current, the magnetic field along the X' axis will be constant, just as in the direct current case in the laboratory frame.
This is the same as moving the coil about the rotating frame coordinate system at the Larmor Frequency. In magnetic resonance, the magnetic field created by the coil passing an alternating current at the Larmor frequency is called the B1 magnetic field. When the alternating current through the coil is turned on and off, it creates a pulsed B1 magnetic field along the X' axis.
The spins respond to this pulse in such a way as to cause the net magnetization vector to rotate about the direction of the applied B1 field. The rotation angle depends on the length of time the field is on, τ, and its magnitude B1.
In our examples, τ will be assumed to be much smaller than T1 and T2.
A 90o pulse is one which rotates the magnetization vector clockwise by 90 degrees about the X' axis.
A 90o pulse rotates the equilibrium magnetization down to the Y' axis.
In the laboratory frame the equilibrium magnetization spirals down around the Z axis to the XY plane.
You can see why the rotating frame of reference is helpful in describing the behavior of magnetization in response to a pulsed magnetic field.
A 180o pulse will rotate the magnetization vector by 180 degrees.
A 180o pulse rotates the equilibrium magnetization down to along the -Z axis.
The net magnetization at any orientation will behave according to the rotation equation. For example, a net magnetization vector along the Y' axis will end up along the -Y' axis when acted upon by a 180o pulse of
B1 along the X' axis.
A net magnetization vector between X' and Y' will end up between X' and -Y' after the application of a
180o pulse of B1
applied along the X' axis.
A rotation matrix (described as a coordinate transformation in #2.6 Chapter 2) can also be used to predict the result of a rotation. Here θ is the rotation angle about the X' axis, [X', Y', Z] is the initial location of the vector, and [X", Y", Z"] the location of the vector after the rotation.
Time varying fields at the Larmor frequency cause transitions between the spin states and hence a change in
MZ. This screen depicts the field at the green hydrogen on the water molecule as it rotates about the external field Bo and a magnetic field from the blue hydrogen.
Note that the field experienced at the green hydrogen is sinusoidal.
There is a distribution of rotation frequencies in a sample of molecules. Only frequencies at the Larmor frequency affect T1. Since the Larmor frequency is proportional to Bo, T1 will therefore vary as a function of magnetic field strength.
In general, T1 is inversely proportional to the density of molecular motions at the Larmor frequency.
The rotation frequency distribution depends on the temperature and viscosity of the solution. Therefore T1 will vary as a function of temperature.
At the Larmor frequency indicated by νo,
T1 (280 K ) < T1 (340 K). The temperature of the human body does not vary by enough to cause a significant influence on T1. The viscosity does however vary significantly from tissue to tissue and influences T1 as is seen in the following molecular motion plot.
Fluctuating fields which perturb the energy levels of the spin states cause the transverse magnetization to dephase. This can be seen by examining the plot of Bo experienced by the red hydrogens on the following water molecule.
The number of molecular motions less than and equal to the Larmor frequency is inversely proportional to T2.
In general, relaxation times get longer as Bo increases because there are fewer relaxation-causing frequency components present in the random motions of the molecules.
The bidirectional arrow indicates that the exchange reaction is reversible.
The energy difference between the upper and lower energy states of A and of B must be the same for spin exchange to occur. On a microscopic scale, the spin in the upper energy state (B) is emitting a photon which is being absorbed by the spin in the lower energy state (A). Therefore, B ends up in the lower energy state and A in the upper state.
Spin exchange will not affect T1 but will affect T2. T1 is not effected because the distribution of spins between the upper and lower states is not changed. T2 will be affected because phase coherence of the transverse magnetization is lost during exchange.
Another form of exchange is called chemical exchange. In chemical exchange, the A and B nuclei are from different molecules. Consider the chemical exchange between water and ethanol.
Here the B hydrogen of water ends up on ethanol, and the A hydrogen on ethanol ends up on water in the forward reaction. There are four senarios for the nuclear spin, represented by the four equations.
Chemical exchange will affect both T1 and T2. T1 is now affected because energy is transferred from one nucleus to another. For example, if there are more nuclei in the upper state of A, and a normal Boltzmann distribution in B, exchange will force the excess energy from A into B. The effect will make T1 appear smaller. T2 is effected because phase coherence of the transverse magnetization is not preserved during chemical exchange.
Copyright © 1997-2017 J.P. Hornak.
All Rights Reserved.