Nuclear Magnetic Resonance Microscopy

Stefanie C. VanGorden


Introduction

Nuclear magnetic resonance (NMR) microscopy is magnetic resonance imaging (MRI) on a small scale (3).   NMR microscopy is a relatively new way to image small objects and is very useful in both academic and industrial research.  NMR microscopy is used by the food and agricultural industries as a tool for analyzing the amount and effect of water in their products. Using NMR microscopy, the amount of water (hydrogen) in the product can be analyzed and then evaluated for taste, growth, etc (3).  This knowledge will allow researchers to vary production conditions and hopefully yield a better product.  The broad, long-term objective of this research is to develop NMR microscopy at the Rochester Institute of Technology (RIT) to a point where it is potentially useful to industry, especially analyzing food for water content.

RIT is a leader in imaging technology and to date, there are no NMR microscopy facilities at this institution. Currently, a Bruker 300 MHz NMR spectrometer is used to obtain spectra.  Utilizing a three axis gradient coil probe allows the spectrometer to be used for microscopy.  The main goal of this research is to document features such as slice thickness and in-plane resolution on the NMR microscope.  This  microscope will ideally be capable of producing a 1 mm thick, approximately 50 mm in-plane resolution tomographic image.
 

Background


Nuclear magnetic resonance imaging was developed in the early 1970's and was largely used for medical research on the human body (3). In 1986, MRI was done on a much smaller scale using NMR spectrometers essentially creating the field of NMR microscopy (3)(10)(13)(8)(12). NMR has widespread applications in gathering information about microscopic properties of biological tissues, pharmaceutical products, and agricultural products (9).

Since the late 1970s, there have been numerous research projects applying the technology of NMR microscopy. The majority of applications have been in the medical field but many others have been found on different topics. Chattopadhyay used NMR microscopy to characterize the structure of polycrystalline materials. Others did research on time-independent point-spread functions and high-gradient NMR microscopy (12)(13). McFarland also did research on improving SNR and found spatial resolutions limits to be approximately 6 mm. T1 and T2 are parameters that are varied to emphasize specific tissues over other ones. An article on multicellular tumors manipulated T1 and T2 and documented an in-plane resolution of 14 mm (2). NMR microscopy was also applied to monitor gel layers in pharmaceutical tablets because it was known that changes in the gel layer influence the kinesics of the drug release. This research used an in-plane resolution of 70 mm and a slice thickness of 650 mm (1). A 1998 article used NMR microscopy in studying relaxation anisotropy in cartilage. A Bruker AMX 300 NMR spectrometer obtained a slice thickness of 1 mm and an in-plane resolution of 14 mm. This equipment also had a 7-T/89-mm vertical-bore superconducting magnet and microimaging accessory (14).

NMR microscopy utilizes a property found certain nuclei called spin. The governing equation for both MRI and NMR is

n = g B     [1]

where B is the magnetic field strength, and g is the gyromagnetic ratio. The gyromagnetic ratio for hydrogen is 42.58 MHz/T. The frequency n is proportional to the gyromagnetic ratio and also the strength of the magnetic field. Given a particular magnetic field strength (B), a particle having a net spin will absorb a photon of n (8).

In NMR microscopy, a magnetic field is applied to a sample containing water (for example) and the property of spin causes protons (Hydrogen) in the water molecules to absorb a photon of frequency na. With currently available NMR magnets na is in the range of 60 to 800 MHz (8). Absorption causes the proton to be excited to the next energy level and all of the excited protons combined can be thought of as the net magnetization. The magnetic field is conventionally placed in the z direction. At equilibrium, the net magnetization is zero Mo and when the system is not at equilibrium the net magnetization is Mz. The equation governing the two relationships is

Mz = Mo ( 1-e-t/T1)     [2]

where T1  is the spin-lattice relaxation time and represents the time needed to change Mz by a factor of e.  T2 is the spin-spin relaxation time and it is always less that T1. The equation governing T2 is

Mxy = Mxyo e-t/T2     [3]

and this is the time needed to reduce the transverse magnetization by a factor of e.  In order to determine the location of protons in the sample, pulsed magnetic fields of 90 degrees or 180 degrees are used. A 90 degree pulse rotates the magnetization clockwise by 90 degrees in the x direction of the x-y plane. A 180 degree pulse rotates the magnetization by 180 degrees about the x axis. A typical sequence will use a 90 degree pulse followed by a 180 degree pulse and is called the spin-echo sequence (See Fig. 1).

Figure 1: Timing diagram for a spin-echo sequence (7).

Fig 1. is just a schematic diagram of a spin-echo sequence for spectroscopy.  It is necessary to add two gradients to perform tomographic imaging.  With the additions of three gradients, this sequence is what allows protons in a sample to be mapped to their x-y locations.  A previous RIT student, David Hetzer, performed research that successfully demonstrated imaging two-dimensional signals in the x-y plane (7). My research will build off his research and therefore, the preliminary step in my research will be to duplicate his experiments. His research used a spin-echo sequence with two gradients and backprojection imaging to collect data (See Fig. Fig 2).
 
 

Figure 2: Example of backprojection imaging(8).

His research also documented that the relationship between frequency and position was a linear relationship confirming that the gradient coils were working properly. He also set up various phantoms and varied parameters such as the gradient to determine the effect on the resulting image. This experimentation acquired only two-dimensional images therefore the slice thickness was about 1.5 cm. All of the information was averaged together in the z direction.

For tomographic imaging, the slice thickness (Thk) is specified and the signal is squished the width equal to Thk. Using the governing equation
n=g Bo, and substituting for Bo we yield

Dn=g Thk Gi      [4]

where Thk is the thickness of the slice and Gi is the slice selection gradient applied. By rearranging, we yield the thickness as a function of the change in frequency (Dn) and inversely proportional to the slice selection gradient.

Thk = Dn/(g Gi)     [5]

Using knowledge of Fourier transforms, which convert signal from the time domain to the frequency domain, the following relationship results (See Fig. 5):

Dn = 2/ tw     [6]

By further graphic interpretation, we can conclude that Dn = 2/tw and substituting this into Eq. 5 yields

Thk = 2/(g Gi tw)     [7]

which will allow us to calculate various parameters. We know that the gyromagnetic ratio is 42.58 Hz/T for hydrogen, Gi is less than 50 x 10-4 T/cm, and tw is 10-6 seconds. Rearranging the equation and solving for the slice selection gradient yields Gi =  500 x 10-4 T/cm which causes a problem because Gi is greater than the maximum gradient of the system by a factor of ten. To compensate for this problem, we will have to utilize the equation

q = 2p g tw B1     [8]

where q is equal to 90 degrees, tw is the amount of time the field is on, and B1 is specified in Figure 5 below. We will make B1 larger which will make tw smaller and plugging into Eq. will yield a smaller Gi. This should yield a slice thickness of 1 mm as desired.

Figure 3: (A) Shows the relationship the result of a Fourier Transform
(B) Illustrates the relationship as a result of the FT.

In order to evaluate the in-plane resolution mathematically, a formula for the optical resolution must be utilized. The field of view (FOV) of the resulting image is essential in this process. The FOV is equal to the frequency of sampling fs, divided by the gyromagnetic ratio g multiplied by the frequency encoding gradient Gf by definition.

FOV = fs / g Gf     [9]

 

Substituting into the above equation yields the optical resolution. The optical resolution (OR) is equal to the FOV multiplied by the product of inverse of pi multiplied by the combined time constant T2* and the quantity divided by the gyromagnetic ratio g times the frequency gradient applied in the x direction Gnx.

OR = (FOV (pT2*)-1)/(g Gnx)     [10]

These variables are located in Fig. 4 for further clarification.
 
 

Figure 4: Illustration of the field of view and T2*.

The result of optical resolution calculated by Eq. 10 is the limit to how small the theoretical resolution will become. Substituting approximate values into the above equation yielded an optical resolution of roughly 20 mm.

Furthermore, the main goal of my research project is to produce one millimeter thick, 50 mm in-plane resolution Magnetic Resonance images using the Bruker 300 MHz DRX spectrometer. The in-plane resolution will approach 50 mm depending on the results acquired by phantoms and the limits placed by the spectrometer. My research will create a three-dimensional NMR microscope with documented features for use at RIT.
 

Methods

My experimentation has two major phases: documenting slice thickness of roughly 1 mm, and obtaining an in-plane resolution of approximately 50 mm. There are no human subjects required for these experiments. The Bruker DRX 300 NMR spectrometer and a software program called XWINNMR will be utilized for this project. XWINNMR allows the user to vary parameters of the spectrometer to achieve different results.

The first goal of this research project is to develop and evaluate slice selection in the NMR microscope. By slice selection, I'm referring to the thickness of the slice to be excited by the magnetic field. All other areas in the sample tube will remain unchanged. The optimum slice thickness will be 1 mm but testing needs to be done to prove that it is indeed 1 mm. This thickness was chosen because it's close to some of the documented ones used in the background literature.  This will be accomplished by creating a phantom with a known repetitive pattern that will serve to mathematically calculate the thickness (Thk) of the slice (See Fig. 5).
 
 

Figure 5: Illustration of slice thickness.

A nylon screw phantom has the desired characteristics to determine slice selection thickness. The diameter of the nylon screw is less than 5mm and it has 32 threads per inch (TPI) which is about 1.2 threads per millimeter. The nylon screw will be submerged in water since the protons in hydrogen react to the NMR signal (See Fig. 6).
 
 

Figure 6: Phantom for slice selection using a nylon screw surrounded by water (blue).

 

The nylon screw phantom will be used to evaluate whether the theoretical calculations did yield a slice thickness of 1 mm. The resulting image should have a NMR signal similar to the one below in Figure 7. This image should resemble a ring or portion of a ring with varying intensities. Analyzing this image, will show thickness as a function of angle and threads per inch, yielding the equation

Thk = f/360 * 1/TPI * 25.4 mm/In     [11]

from trigonometry and converting TPI to threads per millimeter. Below is a Figure illustrating these concepts.
 
 

Figure 7: Resulting image of the slice selection phantom
used to determine the thickness of the slice.

Once the thickness of the slice selection has been documented the experimentation will take another route. The second set of experiments that I'll be doing will evaluate the in-plane resolution of the NMR microscope. Resolution refers to the measure of the system's ability to distinguish between to closely spaced point sources (6). The fundamental limit to this system is that as size of the volume element gets smaller, its signal contribution is reduced and a reduction occurs in the signal-to-noise ratio (3). This phenomenon can be reduced using signal averaging. This will not be concentrated on in this experimentation though.

The theoretical resolution limit of this system is about 20 mms.  To quantitatively test this limit, a specific phantom will be created. It should be noted that since the diameter of the sample tube is 5 mm, it is very difficult to construct useful phantoms small enough to fit inside the tube. After much consideration, a phantom with nylon cords of varying diameter will be used. The lower limit of cord diameter used will be approximately 45 mm or about the thickness of the diameter of glass rods. The upper limit will be approximately 139 mm for fishing line. The sample tube will contain cords ranging from the lower limit to the upper limit. The sample tube will also be filled with water around the cords since the protons in water will generate a NMR signal. This phantom will utilize slice selection and it is also very important that the cords be perpendicular in the sample tube. An illustration of the sample phantom is located in Figure 8 below.
 
 

Figure 8: Illustration of phantom to evaluate in-plane resolution.

The goal of this phantom is to use the known diameter of the nylon cords and the resulting NMR image to determine the in-plane resolution of the system. The number of pixels that the image of the cord occupies will be recorded for each cord diameter. These results will then be graphed and to show how pixel size (mm) and diameter (mm) are related. The slope of this curve should equal one for the ideal case where the objects are the same size in the image as they were when measured. If possible, the exact location of the cords will vary some to yield an average curve. This is attempting eliminate samples taking up more pixels because of falling half on one pixel and half on the next pixel. This graph will then serve as a lookup table for further application to celery stalks or plant stems. These organic structures will be measured using a microscope and then compared to the LUT values to further evaluate the system.
 
 

Bruker Equipment:

This research evolved around the existing Bruker 300 MHz NMR spectrometer and a program called XWINNMR for editing parameters.  A portion of the research time was spent learning how to operate the spectrometer, and use the software.

There are two probe heads on the spectrometer, one for taking spectra (qnp #7) and one for imaging (bbi #11).  This spectrometer is normally used for taking spectra so the probe head was changed and replaced for each day of data collection.

Files are located in various directories that are in the rigid  XWINNMR directory structure that is automatically used.  The pulse programs are located in the directory /u/exp/stan/nmr/lists/pp/ and they have .ssse extensions.  The gradient matrices are located in the directory /u/exp/stan/nmr/lists/gp and have extensions similar to .ssse.r.  Images are located in /disk2/data/scv996/nmr/.  The pulse programs and gradient matrices are located in the appendix.

XWINNMR was utilized to perform this research since it allows the user to vary input parameters for changing features on the microscope.  The table below contains many of the commands that were used.  Research was done in manual control mode that turns off the sweeping feature of the spectrometer.
 

Table 1:  Common XWINNMR commands and their meanings.

XWINNMR Command

Description

rsh

Read in shim file for a particular probe

wsh

Write a shim file for a particular probe

edhead

Change probe seen by the software

acq

Shows the FID (Free induction decay) window

pulsdisp

Allows the user to view a simulation or timing diagram

rga

Receiver gain adjustment

wobb

Allows user to tune the probe

halt

Ends the wobb command

zg

Runs the program specified in the eda display

eda and edp

Used to change programs or length of pulses

xfb

Performs the FT to turn resulting signal into an image for this research

NS

Number of scans taken

DS

Number of dummy scans

P1

Duration of the 90 degree pulse

P2

Duration of the 10 degree pulse

PL1

Amplitude of the above pulses

D2

 

L1

Loop counter parameters (choose 4 and it'll perform 4*256 scans)

The time it takes for image collection varies.  The user can specify the image size and the number of repetitions.  Images collected were 256x256.  It takes about 1.5 minutes to obtain an image.  Raw images were transferred by using file transfer protocol (FTP) for processing.  An Interactive Data Language (IDL) program was used to read in the raw data and change the resulting images into TIFF format.  This IDL program is located in the appendix.
 

Results and Discussion


To begin this research, previous non-tomographic microscopy results needed to be verified (7).  This research essentially created a NMR microscope which would take the hydrogen signal from the entire sample tube range of  1.5 cm and average it to create an image. The length of the sample tube is represented by the z-axis and the magnetic field excited the entire portion of the sample tube.  The resulting image has all the sample values in the z-direction averaged to yield one z value for each x-y location.  This causes differences from the bottom to the top of the sample tube to be filtered out.  The timing diagram for this pulse sequence is located in Figure 9.  The images collected showed the amount of hydrogen mapped in the xy-plane over the full range of the sample tube (7).

Figure 9:  Diagram illustrating the two-dimensional sequence.

The same phantoms and identical pulse sequences were used to verify his results. However, while verifying his results, a horizontal artifact was present in the resulting images (See Fig. 10).  This image is of capillary tubes filled with water. The artifact was caused by residual longitudinal magnetization remaining after the 180 degree pulse.  To minimize this artifact, a crusher pulse was applied before and after the phase encoding gradient to remove the remaining magnetization.  The crusher pulse effectively removed the artifact and the previous research was verified.

Figure 10: Image with a horizontal artifact present.










Slice Thickness Documentation

Nylon Screw Phantom

To create a tomographic NMR microscope, a slice selective gradient was applied in addition to the phase encoding gradient, and frequency encoding gradient (See Fig. 11).  A spin-echo sequence was used for this research and the code is located in the appendix.   This sequence, by manipulating the pulse width and the values for the slice selection gradient, caused a specific portion of the sample tube to be excited and created a tomographic NMR microscope.  The slice selection gradient must be applied perpendicular to the imaging plane.  The phase encoding and frequency encoding gradients need to be applied in the same plane as the slice but orthogonal to each other (See Fig. 12).  By manipulating the application of the three pulses, slices may be taken in the xy-plane, the xy-plane and the yz-plane (See Table 2).  Oblique imaging can also be implemented by applying combinations of the various gradients to yield the desired angles.  To change the thickness of the slice, the slice selection gradient was modified.  The thickness of the slice is found using the formula Thk = Dn/(g Gs) where the thickness is inversely proportional to the slice selection gradient.  Gradient values implemented into the time sequence are percentages.  The maximum gradient is 50 G/cm for this spectrometer and by using a slice selection gradient of 100%, the gradient applied will be 50 G/cm.  Using a value of 10% would yield a gradient of 5 G/cm.  Since thickness is inversely proportional to the slice selection gradient, implementing a value of 10% will yield a thicker slice than that of 100%.
 
 

Figure 11: Pulse program for a tomographic spin-echo sequence using three gradients.

Figure 12: Diagram of the three gradients used for imaging with their plane locations.

Table 2: Gradients for imaging in various Planes.

To evaluate the performance of the tomographic sequence, a nylon screw phantom was imaged to attempt to document slice thickness.  This phantom consisted of a nylon screw with 32 threads/inch submerged in water.  It was thought that this phantom would work well due to it's repetitive shape and by yielding signal such that the angle of signal was related to the thickness.  Images were taken with various gradients (See Fig. 13).  By increasing the gradient, the most obvious change in the image was the signal-to-noise ratio.  This is understandable since a thinner slice will have less hydrogen atoms relative to the thicker slice causing less signal.

Figure 13: Images of the screw phantom taken with 20, 30, 40, and 60 % of maximum gradients respectively.

While trying to document the results from this phantom, it became apparent that if multiple turns were present  in the image, there'd be no way to determine how many turns occurred.  Also,  the input for the 90 and 180 degree pulses is square.   Therefore, the Fourier Transform of this input is a sinc function.  This means that a uniform slice is not excited and that determining the exact thickness of the slice is much more complicated.  Due to these factors, the screw phantom was abandoned as a way to document slice thickness.  This phantom did serve to illustrate that by changing the z gradient, the signal-to-noise ratio varied from image to image as seen in Figure 13 above.


Cone Phantom

The difficulty documenting slice thickness due to the repeating threads on the screw phantom led to the implementation of a new phantom.  This phantom consisted of a hollow plastic cone submerged in water.  The cone was originally used for.....????   The premise behind this phantom was to vary the gradients (thickness)  to determine if the resulting images reflected a thicker or thinner circle from the cone.  The angle of the cone was found to be degrees and the width of the cone wall was found to be 0.55 mm (See Fig. 14).   Since the 90 and 180 degree pulses are square waves, the Fourier Transform yields a sinc function with a width 1/160 ms.  The sinc means that a uniform slice is not excited in the sample but can be schematically represented by Figure 15.

Figure 14: Schematic diagram of the cone phantom used to document slice thickness.

Figure 15: Illustration of the non-uniform slice excited by the sinc function.

Utilizing information about the sinc function (See Fig. 15), showed the relationship Dn=2/tw.  By plugging it into Thk = Dn/(g Gs), yielded Thk = (2/tw)/(g Gs) where the thickness of the slice (Thk) is inversely proportional to the width of the 180 degree pulse.   The 180 pulse was 160 ms in duration. By calculations, this yielded a slice thickness of 0.01 mm (gradient = 60%) and slice thickness of 0.59 mm  (gradient=10%).  The original goal was to produce a slice equivalent to 1 mm.  Due to this optimal goal, the above equation was rearranged to solve for the width of the pulse that would create a 1 mm slice.  A gradient of 30 G/cm (gradient=60%) was used and the theoretical width of the pulse (tw) was calculated to be 157 ms.  Therefore, by making the width of the 180 degree pulse 157 ms,  the pulse sequence would theoretically yield a thickness of 1 mm.  The width of the 90 degree pulse is 7.5 m s or half that of the 180 degree pulse.  A gradient of 60% was used so that the thickness of the slice could increase or decrease without changing the width of the RF pulses.  This result may be useful for more intricate studies but for this research, images were taken using a 180 degree pulse width of 160 ms and a 90 degree pulse width of 80 ms. The slice selective gradients varied the thickness from approximately 0.01-0.59 mm.  These pulse widths were used for all of the collected images and thicknesses seemed satisfactory to see changes in the images.  For this piece of equipment, as mentioned previously, the maximum gradient is 50 G/cm but this number may not be exact although the theoretical calculations assume that it is.  It should also be mentioned again that the pulses were square so a non-uniform thickness was excited in the sample causing the signal to be generated from an average of the thickness that followed a sinc pattern.  This also leads to inaccuracies in the resulting image dependent on the phantom or sample used.

Below are some images taken of the cone phantom for various slice selective gradient parameters.  Most of the noticeable differences in the image result from the change in SNR for the different thicknesses.  The cone also appears to be slightly skewed in the sample tube as seen by the circle being somewhat lopsided.

Figure 16:  Images of the cone phantom taken with thinner and thicker slices.

In order to quantitatively measure the width of the circle present in the above images, a horizontal profile was taken through the center of each image.  The horizontal profiles are located below in Figure 17.  For these profiles, the x-axis represents the width of the image which is 256 pixels and the y-axis is the NMR signal created by hydrogen present in the water molecules of the excited slice.  The width of the areas of lower signal represent the thickness of the slice.  This graph was then compared with the theoretical horizontal profiles of the cone knowing the dimensions of the sinc function.
 
 

Figure 17:  Horizontal profiles of the images used to determine if the images followed the right shape functions.

The top image in Figure 17 shows the horizontal profile for a thick slice and the lower image shows an image of thinner slice. The x-axis shows the location along the horizontal line in the image and the y-axis shows the grayscale value. The shape of these profiles matches the expected relationship. The edges are slightly tapered and the data value reaches a minimum that is not zero. Visible in the plots is the difference in the amount of signal-to-noise from varying the slice thickness. These slices vary from about 1 mm (Left) to a thin slice of ~0.3 mm (Right). These are theoretical values. Due to the small slice thickness relative to the geometry of the cone, there is not much difference in the width of the non-signal areas for the different thicknesses. These results make sense when conceptualizing the cone phantom. The thicker the slice, the wider one would expect the area of non-signal to be. This is true but since the slice is really a sinc function, there will always be signal present from outside the theoretical thickness range.



In-Plane Resolution Documentation

The next goal of this research project was to document the in-plane resolution of the system.  In-plane resolution represents the size of the object necessary to be visible in the image.  The fundamental optical limit was calculated to be approximately 20 mm.  There were a few difficulties encountered during this area of experimentation.  In order for the phantom to be relatively accurate, it needed to be very small in diameter and also have a constant shape so that minimal effects would occur when the signal was averaged over the thickness (~0.5 mm).  Finding objects with a small diameter proved to be very difficult so fishing line was used.  Human hair is around the same diameter but the line is much tougher.  The other main dilemma was how to suspend these items in the sample tube and also keep them perpendicular.  Fishing line also sometimes floats in water to add to the burden. After trial and error, the best method to keep the line perpendicular to the imaging plane (xy-plane) was to place the line inside capillary tubes (See Fig. 18).  A micrometer was used to measure varying diameters of fishing line.  The fishing line varied in diameter from 87-139 mm and the width of the capillary wall was measured to be 4.25 mm.  The line and capillary tubes were then submerged in water to complete the phantom.  All of the line was easily visible in the images as seen in Figure 19.

Figure 18: Skematic diagram illustrating fishing line phantom design.

Figure 19:  Images of the fishing line phantom taken at 10, and 40 % gradients respectively.

Since this really was not a limiting resolution for the system, glass rods were stretched to diameters of ~50 mm and suspended in water.  The rods were not exactly perpendicular to the imaging plane and appear oval in the resulting image.  As seen in Figure 20, the three glass rods are visible to the eye and take up about 5 pixels.

Figure 20:  Image of the glass rod phantom taken with 10 % gradient.

No other phantoms were used for in-plane resolution documentation due to the difficulty in finding phantoms with diameters small enough to be useful.  Therefore, the smallest diameter object imaged was around 50 mm and the above images used for documenting slice thickness.

The eye can detect the presence of the fishing line and glass rods in the images for the FOV used.  This however, doesn't give us an accurate portrayal of the relative size of these objects. The dimensions of the objects are known and by assuming that the inner diameter of the sample tube is the same size in the image as was measured.  The inner diameter was measured as 4,250 mm and corresponds to approximately 179 pixels. Dividing these numbers yielded 23.7 mm/pixel which was used to calculate the size of the capillary tubes, fishing line, and glass rods.  This formula is listed below in Eq. 12.

          Image Size = (23.7 mm/pixel) * (Number of pixels in image)     [12]

The calculated image size numbers should ideally match the measured dimensions of the objects and yield a slope of 1 for the graph.  This was not the case.  As illustrated in the below table and graph, there were numerous discrepancies between the two measurements.

Table 3: Calculated values for measured diameter [m m] and  image diameter [m m].

Object Name

Object Size [m m]

Image Size [m m]

Glass Rods

45

118

Small Line

87

248

Medium Line

126

379

Large Line

139

426

Inside dia. Cap. Tube

1000

900

Outside dia. Cap. Tube

1500

1398

Inside dia. Sample Tube

4250

4250




Figure 21: Graphs of measured diameter [m m] vs. calculated image diameter [m m].


 

For the smaller objects, the image size is more than double that of the original size but for the larger measurements, the size in the image is smaller than the measured value.  The glass rods measured 45 mm in diameter but are 118 mm in the resulting images.  Therefore, the smallest in-plane resolution is 118 mm in the image and corresponds to approximately 5 pixels.  The cause of this discrepancy in size is unclear.  It might be due to rods not being perpendicular in the sample tube.  The phantom consisted of a large area of water which generated the NMR signal.  Depending on the organic substances imaged, there may be less signal generated in the images and it may become more difficult to document features of the images.

Errors were present in the images due to the fishing line or glass rods not being exactly perpendicular to the imaging plane.  Also, this documentation used a fixed FOV which allowed the entire diameter of the sample tube to be present in the images.  Modifying the FOV such that only one capillary tube is present in the image would likely increase the in-plane resolution. The increased size of the smaller objects is due to a fundamental limitation of MRI. This causes the objects to become smeared in the images and the exact reason is unknown. Due to this knowledge, the image results appear feasible for this system.
 
 


Organic Applications

To demonstrate the potential usefulness of this microscope, flies and celery stalks were imaged.  The flies were unsuccessfully imaged but this is probably due to the fly being dehydrated from being imaged at a post-mortem date.  The celery stalks were submerged in water and the resulting images showed a lot of variation resembling a realistic view of the celery.  The signal variation represents areas of high hydrogen content (white) and low hydrogen reaction (black).  Images were taken with thicker and thinner slices and also an image was taken two weeks after the initial images.  Structure variations are present between all images.  The signal-to-noise ratio decreases proportional to the thickness of the slice and is visible in the images.  Decomposition is present in the right hand image and the structures appear to have broken down also.  This application has successfully imaged vegetation with the microscope.  Future work could be done on germinating seeds and NMR images could potentially capture various stages of development.

Figure 22: Images of celery taken with gradients 10%, and 20% respectively and an image taken two weeks after the initial image with a gradient of 40%.

Conclusions

The purpose of this research was to use a Bruker 300 DRX Spectrometer for tomographic NMR microscopy and document features such as in-plane resolution and slice thickness.  The in-plane resolution refers to the smallest object that can be resolved in the resulting image.  Slice thickness for this research represents the amount of sample to get excited by the magnetic field documented in the z-direction.  Ideal goals were a slice thickness of 1 mm and an in-plane resolution of 50 mm.

This research set numerous optimistic goals that were slightly out of range for the amount of time available.  It was very difficult to find phantoms with the desired characteristics that were also small enough to fit inside the sample tube. It was also difficult to determine a slice thickness due to the pulse being shaped like a sinc in Fourier space and a non-uniform slice being excited from the phantom.  Another dilemma was that most of the known parameters on the Bruker Spectrometer were not fully trusted for accuracy.  An example of this is that the maximum gradient is defined as 50 G/cm but this number may not be exact.  This value changes the theoretical calculations if inaccurate.

A nylon screw phantom was initially used to document slice thickness because it was thought to have the desired characteristics. The screw was submerged in water to create the nmr signal. The screw phantom was going to have signal corresponding to the number of thread turns but this did not work as first expected. There was no way to discriminate between thread turns of more than one. Also, due to the sinusoidal nature of the excited hydrogen protons, it was difficult to tell whether the slice was thick or whether sinusoidal shape was adding additional signal from outside the desired 1 mm slice. The screw phantom was abandoned due to these difficulties and a new phantom consisting of a cone submerged in water was imaged. Intricate measurements were taken of this phantom and the theoretical amount of signal was modeled for use in determining the slice thickness. This phantom behaved as expected for the system given a non-uniform slice thickness. The horizontal plots of the images show both signal-to-noise changes between the images and also a relative shape that takes into account the slice thickness.

The next goal of this research was to document the in-plane resolution using a constant field of view. The major difficulties with this phantom were that it was difficult to find objects with the desired characteristics that were small enough to test the in-plane resolution limits. Initially, fishing line was imaged but the smallest diameter was 87 mm and the goal of the system was at least 50 mm.  To test the limits further, glass rods were stretched and submerged in water since they were the smallest objects available for use with diameters of approximately 45 mm. Glass rods are visible in the resulting images.  It was very difficult to keep the glass rods perpendicular in the sample tube and this is visible in the images by their oval shape. Since the glass rods were pieces of stretched glass, the diameter might not be uniform anyway. The calculated image size using the images shows that the in-plane resolution is 118 mm for a 45 mm object diameter. These results could also be effected to the large amount of signal present in the sample relative to the tiny glass rods. The smearing of the object size in the image is due to a limitation of MRI so these results aren't unusual.

Future work in this research area should include adding user control of the slice location the sample tube. Some ideas for this are in the following paragraphs.

Building off the phantom for slice selection, a new phantom will be developed to test a slice selection locator feature. Assuming that that the results from the first phantom adequately documented the thickness of the slice, this feature will allow the user to freely choose the particular location of the 1 mm slice. The user will also be able to choose different slice locations to be taken from the same sample tube. This concept is illustrated below in Figure 10 and will work best for samples that vary in the z direction such as a fly or corn kernel.

Figure 23: Fly with varying signal components in the z direction
and illustration of the slice selection locator.



Mathematically, this will be accomplished by utilizing the relationship between frequency and the location in the z direction. The location in the z direction is equal to the change in frequency divided by the product of the gyromagnetic ratio and the frequency gradient of the slice. Change in frequency will be equal to the frequency we apply minus a reference frequency at zero.

z = Dn/(g Gslice) [13]

Dn = n - no [14]

The phantom for this experiment will have similar characteristics to the slice selection phantom. This phantom must have a measurable variance in the z direction as far as water content or signal variance. A simple phantom can be constructed with capillary tubes filled with varying amounts of water in them over the z range of the detector. This should result in a variance of signal proportional to the z location and give us enough knowledge to narrow down where the slice is being taken. An illustration of this phantom is in Figure 11 below. After getting an approximate region of where the slice is being taken, the nylon screw may be placed in a measured amount of water and the results should yield enough information to determine the final location of the slice.

Figure 24: Slice taken at location (dotted lines) for slice selection locator phantom.

Given that the thickness of the slice and the in-plane resolution are both accurately documented, the phantom mentioned above should allow the specification to the location of the slice. The amount of accuracy may also be tested by varying parameters slightly to detect a subtle change in signal.
 
 
 