Fluorescent Database - Sergio Gonzalez

This page was design with the purpose of making available bispectral measurements for anyone who wants to analyze them. The database is made up of sampling different printing processes under normal reproduction conditions of solid colors and a variety of fluorescent materials. Throughout the experiment a bispectral spectrophotometer (BFC-450) manufactured by Labsphere was used to measure the samples

**Links to the individual bispectral files are located in the tables were the samples are listed.
All of the data files can be downloaded here. This is a 6.5MB zipped file. It may take a long time to download, and occasional connection failures may occur. The relevant thesis can also be downloaded: SergioThesis.pdf. This is a 2.8MB PDF file.

Below there is brief discription of bispectral measurement and how to obtain colorimetric values out of the files.

Bispectral Method

Bispectral spectrophotometric instruments can make colorimetric measurements by taking into account the contribution of both the fluorescent and the reflected component to the total radiance of a sample. For the bispectral method one monochromator is located between the instrument light source and the sample to be measured. The function of the monochromator is to separate the radiation from the instrument's light source into its spectral components before it reaches the sample. The second monochromator is located between the sample and the photodetector, which separates the radiation leaving the sample surface into its spectral components.

 

This section is devoted to describing the process in which colorimetric values are obtained (CIELAB) from the spectral radiance factor of a fluorescent sample. Two arbitrary samples were chosen to exemplify how the values are calculated. The samples chosen were a fluorescent orange golf ball and a green fluorescent plastic sample.

The work in this investigation was done in a bispectral spectrophotometer (which gives data of complete spectrum of light emitted by the sample at each wavelength of irradiation). The procedure will have as starting point the output data from bispectral measurements.

The instrument output is a matrix with wavelength by wavelength contribution of light excitation and emission. The columns in Figure 5 correspond to the excitation while the rows correspond to the emission wavelengths, the values within the diagonal correspond to the reflected component while the values off-diagonal correspond to the fluorescent contribution. Figure 6 shows the graphical representation of the matrix form. The xy plane corresponds to the excitation and emission wavelengths while the z-axis represents the radiance factor.

 

Figure 5 Part of Matrix of a bispectral measurement from a green fluorescent sample


Figure 6. 3D representation of a bispectral measurement from a green fluorescent sample.

 

After having the matrix representation of a fluorescent sample Equation 1 can be rewritten into Equation 2, where the total radiance factor is in terms of both the emission wavelengths (irradiating light) and the excitation wavelengths (light coming out of the sample) as well as the reflected radiance factor and fluorescent radiance factors. Now the total radiance will be describe with the symbol "b" instead of the "I", previously used because the total radiance now is a function depending on two variables.

Tristimulus Values

The calculation of the tristimulus values starts with the bispectral radiance factor (bT(m,w) matrix form) which is expressed in function of the excitation (m) and the emission (w) wavelengths.

The bT(m,w) is multiplied by the specified light source (FIl(m)) for colorimetric calculations as shown in Equation 6. Then the resultant matrix is summed over the excitation wavelength to obtain an array, which becomes emission wavelength dependent.

Once obtaining tT(w), which can be called the stimulus function, the XYZ can be obtained with traditional matrix colorimetric approach as shown in Equation 7. In the present work, the CIE 1931 standard colorimetric observer (2o) color matching functions were employed.

Also by using equations 4 and 5 the stimulus function for the reflected and the fluorescent components can be obtained separately. Then by solving Equation 7 the tristimulus values of any stimulus function can be derived. The sum of each tristimulus value from the reflected component with the fluorescent component must equal each tristumulus value of the total radiance factor (example XF+XS=XT).

For a better understanding how the math works an analogy for non-fluorescent samples can be made using bispectral nomenclature. Equation 9 shows the stimulus function for a non-fluorescent sample. In this case, the sum symbol is replaced with the integral symbol because a continuous function replaces the discrete function of the fluorescent sample. Both equations have the same function to sum over the excitation wavelengths. Since there is no fluorescent component the equation is simplified because there is no excitation dependency. Then by applying Equation 7 the XYZs can be derived.

Equation 9. Approach using bispectral terminology for non-fluorescent materials

CIELAB Values

Once the XYZ values are calculated, the CIELAB colorimetric values are obtained using the traditional approach Equations 10 through 12 (for most of the work a D50s light source was employed). In Equations 10 through 12 the subscript n refers to the tristimulus values of a perfectly diffuse reflector.

 

Difference Between Bispectral Measurements and Conventional Instrumental Method

The main difference of the bispectral instruments from the conventional spectrophotometers is the incorporation of two monochromators into the instrument. This way the measurement becomes light-source independent and the full bispectral radiance factor can be obtained in a matrix form as a function of the excited and emitted wavelengths.

 

The types & quantity of samples

On the first stage the analysis was based on: seven prints (paper with color patches of 100% CMYK and 50 % CMYK), one print (paper with patches of 100% CMYK and 40 % CMYK), and one print (paper with patches of 100% CMYK). In total they were 76 measurements. They were measured with the intention to analyze the effect of fluorescent component in color determination.

Among the different printing process used to generate the samples were: two color proofers (3M & Epson), two thermal printers (Kodak XLT 7720 & Fujix Pictography), two RIT Lithographic presses, and a combination of inkjet printers with different quality papers. Table 1 shows the complete list of printed samples used on the first stage of measurements.

Table 1 Sample list for the first stage

KodakXLT_100%_Cyan

3m-conventional film proof 100% Cyan

KodakXLT_100%_Black

3m-conventional film proof 100% Black

KodakXLT_100%_Magenta

3m-conventional film proof 100% Magenta

KodakXLT_100%_Yellow

3m-conventional film proof 100% Yellow

KodakXLT_50%_Cyan

3m-conventional film proof 50% Cyan

KodakXLT_50%_Black

3m-conventional film proof 50% Black

KodakXLT_50%_Magenta

3m-conventional film proof 50% Magenta

KodakXLT_50%_Yellow

3m-conventional film proof 50% Yellow

KodakXLT_White-Paper

3m-conventional film proof white film

Printed Litho Web feed 100% Cyan

Printed Epson digital proof 100% Cyan

Printed Litho Web feed 100% Black

Printed Epson digital proof 100% Black

Printed Litho Web feed 100% Magenta

Printed Epson digital proof 100% Magenta

Printed Litho Web feed 100% Yellow

Printed Epson digital proof 100% Yellow

Printed Litho Web feed 40% Cyan

Printed Epson digital proof White

Printed Litho Web feed 40% Black

Fuji_Pict._100%_Cyan

Printed Litho Web feed 40% Magenta

Fuji_Pict._100%_Black

Printed Litho Web feed 40% Yellow

Fuji_Pict._100%_Magenta

Printed white paper web offset Litho

Fuji_Pict._100%_Yellow

Litho sheet feed 100% Cyan

Fuji_Pict._50%_Cyan

Litho sheet feed 100% Black

Fuji_Pict._50%_Black

Litho sheet feed 100% Magenta

Fuji_Pict._50%_Magenta

Litho sheet feed 100% Yellow

Fuji_Pict._50%_Yellow

Litho sheet feed 50% Cyan

Fuji_Pict._White_paper

Litho sheet feed 50% Black

Hp870cxi_100%_Cyan_hp_paper

Litho sheet feed 50% Magenta

Hp870cxi_100%_Black_hp_paper

Litho sheet feed 50% Yellow

Hp870cxi_100%_Magenta_hp_paper

Litho sheet feed white paper

Hp870cxi_100%_Yellow_hp_paper

Xeror-Cyan-100%-riverside paper

Hp870cxi_50%_Cyan_hp_paper

Xeror-Cyan-50%-riverside paper

Hp870cxi_50%_Black_hp_paper

Xeror-Black-100%-riverside paper

Hp870cxi_50%_Magenta_hp_paper

Xeror-Black-50%-riverside paper

Hp870cxi_50%_Yellow_hp_paper

Xeror-Magenta-100%-riverside paper

Hp870cxi_white_hp_paper

Xeror-Magenta-50%-riverside paper

Hp870cxi_100%_Cyan_riverPaper

Xeror-White-paper-riverside paper

Hp870cxi_100%_Black_riverPaper

Xeror-Yellow-100%-riverside paper

Hp870cxi_100%_Magenta_riverPaper

Xeror-Yellow-50%-riverside paper

Hp870cxi_100%_Yellow_riverPaper

 

Hp870cxi_50%_Cyan_riverPaper

 

Hp870cxi_50%_Black_riverPaper

 

Hp870cxi_50%_Magenta_riverPaper

 

Hp870cxi_50%_Yellow_riverPaper

 

Hp870cxi_white_riverPaper

 

In the second stage 60 samples were measured, which were considered to have fluorescent properties under normal conditions seen by an average person as well as under black light. The list of samples is shown in Table 2. The origin of the samples is broad, it ranges from textiles, plastics, crayons, highlighters to color catalogs, etc. The main purpose of this second stage is to build a small database of fluorescent materials for future study and research.

Table 2 Sample list for the second stage non-printed material

Riverside Array Hyper laser & inkejet multipurpose paper

Textiles

Red

Magenta

Orange

Light Blue

Yellow

Orange

Green

Yellow

Magenta

Magenta un cut

Paint sample ( oxfrod index card)

Plastic film

Colorations paint Red

Orange

Colorations paint Orange

Green

Colorations paint Yellow

White

Colorations paint Green

White plastic (ciba white scale)

Colorations paint Blue

Num 8

Alex poster paint Magenta

Num 9

Alex poster paint Green

Num 10

Sanford highlighter (oxford index cards)

Num 11

Sample

Num 12

Crayola markers ( oxford index cards)

Baked Scupley III polymer clay

Hot pink

Red

Infra Red

Orange

Laser Lemon

Yellow

Hot Magenta

Green

Outrageous Orange

Blue

Unmellow Yellow

Purple

Atomic Tangerine

Radiant color Pigments (hercules)

Electric Lime

Orange

Blizarrd Blue

Pink

Shocking Pink

Blue

Purple Pizzazz

Red

Magic Mint

Chartreuse

Index Card

Magenta

Sample

Orange Yellow

Golf Ball

Orange Red

White

Green

Yellow

Cerise

Orange

3M Scotchlite Retroreflective Sheeting

Macbetch color checker

Orange uniform

Moderate red

Orange grided

 

Extracted from the thesis:

Gonzalez,Sergio "Evaluation of Bispectral Spectrophotometry for Accurate Colorimetry of Printing Materials" ,RIT, Jun 2000, pp 14-23