Types: Transmit & Receive, Receive Only, and Transmit Only
n = resonant frequency
L = Inductance
C = Capacitance
B1 must be perpendicular to Bo.








A laboratory to rotating frame of reference converter.

Inputs: n and no
Outputs: MX'(t) and MY'(t)
Two Input Signals: Cos(A) and Cos(B)
Output Signal: 1/2 Cos(A+B) and 1/2 Cos(A-B)




The specific absorption rate (SAR) is the limiting measure.
| Static Magnetic Field ( Bo ) | |
|---|---|
| Maximum clinical imager magnetic field | 2.0 T |
| Maximum inadvertant exposure | 0.0005 T |
| Rate of Change of Magnetic Field (dB/dt) | ||
|---|---|---|
| System | < 6 T/s | |
| Axial Gradients | t > 120 ms | < 20 T/s |
| 12 ms < t < 10 ms | < 2400/t(ms) T/s | |
| t < 12 ms | < 200 T/s | |
| Transverse Gradients | < 3x above | |
| RF Power Deposition | ||
|---|---|---|
| Temperature (37o C normal) | Change | < 1 o C local |
| Maximum Head | < 38 o C | |
| Maximum Trunk | < 39 o C | |
| Maximum Extremities | < 40 o C | |
| Specific Absorption Rate (SAR) | Whole Body | < 0.4 W/kg |
| Any 1 gram of tissue | < 8.0 W/kg | |
| Head (averaged) | < 3.2 W/kg | |
| Acoustic Noise Level | |
|---|---|
| LPeak | 200 Pascals |
| Leq. averaged over 1 Hr | 105 dBA |